ICCNexergy

Leaders in Rechargeable Power Systems

mw40-elpac-power-supplies

MW40 Series

40 Watt Medical Desktop Power Supply

  • High Efficiency: Level V
  • Fully Regulated DC Output
  • Lifetime Expectation >5 years
  • Hold-up Time >16ms at full load
  • Medical Approval - EN60601-1 Class I
  • TUV-RL
  • RoHS
  • CE
  • cURus
Print Datasheet
Elpac Part NumberOutput VoltageOutput CurrentPeak Current1Total Regulation2Typical Efficiency3
MW4009-760F-NC-BK9.0V4.4A5.3A±5%84%
MW4012-760F-NC-BK12.0V3.3A4.0A±5%86%
MW4015-760F-NC-BK15.0v2.6A3.2A±5%87%
MW4018-760F-NC-BK18.0V2.2A2.6A±5%87%
MW4024-760F-NC-BK24.0V1.7A2.0A±5%88%
  • Notes
  • 1 Maximum peak load (48W) lasting 500ms with a maximum 10% duty cycle.
  • 2 Includes initial setting, line regulation, load, regulation, and thermal drift.
  • 3 Typical at 115VAC
Input
Input Voltage 85 – 264VAC 100 – 240VAC Nominal
Input Frequency 47 – 63Hz
Input Current <1A rms
Inrush Current <37A at 230VAC cold start
Zero Load Power Consumption 0.3W
Touch Leakage Current
<50µA @ 132VAC @ 60Hz
<75µA @ 264VAC @ 60Hz
Output
Output Voltage See Table
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <.25s
Hold-Up Time >16ms
Ripple & Noise <1% pk-pk ** *
Over Voltage Protection 110-135%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protetion 120 - 180%
Short Circuit Protection shutdown, auto-restart (hiccup mode)
  • Notes
  • * Ripple and noise measured with 20MHz bandwidth; 10μF tantalum capacitor in parallel with a 0.1μF ceramic capacitor.
General
Efficiency Avg Efficiency 86.4% @ 115VAC; 85.7% @ 230VAC
MTBF min. 200,000 hours demonstrated
Size 5.0 (127.0mm) x 3.0 (76.2 mm) x 1.3 (33.0mm)
Weight 0.8 lbs (.36 kg)
Environmental
Operating Temperature 0 - 40C, linearly derated to 50% load at 60C
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by TUV to the following:
cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90
CB per IEC60601-1
CE marked to LVD
Input Configuration
Standard Input Cable Not Provided
Connection on Power Supply Body IEC 320 C14 Receptacle
Output Configuration
Standard Output Cable 6 ft.
Cord Size 2x16awg
Connector (PSU side) Switchcraft 760 or equivalent
Mating Connector Switchcraft 712A or equivalent
Output Pin Assignments
Center +v1
Outside Return
mechanical drawing
ICCNexergy has the capability to modify our standard power supplies to meet custom specifications. With a broad standard power platform, we can quickly deliver a modified power solution in as little as three weeks. Modified options include:
  • Grounded Output
  • Custom Connectors
  • Ferrite Cores
  • Custom Imprinting
  • Special Packaging/Labeling
  • Overmolding
  • Output Cord Lengths
  • White Case

Inventory Results For MW40

Part Number Distributor Quantity Inv. Date
MW4009-760F-NC-BK NEWARK/ELEMENT14 90 2/3/2012
MW4009-760F-NC-BK ALLIED ELECTRONICS, INC. 15 12/30/2011
MW4009-760F-NC-WH ALLIED ELECTRONICS, INC. 1 12/30/2011
MW4009760FNCBK PEI-GENESIS 14 2/3/2012
MW4012-760-NC-WH ALLIED ELECTRONICS, INC. 45 12/30/2011
MW4012-760F-NC-BK DIGI-KEY CORPORATION 501 2/3/2012
MW4012-760F-NC-BK MOUSER ELECTRONICS 62 2/3/2012
MW4012-760F-NC-BK ALLIED ELECTRONICS, INC. 263 12/30/2011
MW4012-760F-NC-WH DIGI-KEY CORPORATION 100 2/3/2012
MW4012760FNCBK PEI-GENESIS 529 2/3/2012
MW4015-760-NC-BK ALLIED ELECTRONICS, INC. 196 12/30/2011
MW4015-760-NC-WH ALLIED ELECTRONICS, INC. 3 12/30/2011
MW4015-760F-NC-WH DIGI-KEY CORPORATION 284 2/3/2012
MW4015-760F-NC-WH MOUSER ELECTRONICS 54 2/3/2012
MW4015760FNCBK PEI-GENESIS 38 2/3/2012
MW4015760NCWH PEI-GENESIS 1 2/3/2012
MW4018-760F-NC-WH DIGI-KEY CORPORATION 6 2/3/2012
MW4024-760F-NC-BK MOUSER ELECTRONICS 9 2/3/2012
MW4024-760F-NC-BK DIGI-KEY CORPORATION 7 2/3/2012
MW4024-760F-NC-BK ALLIED ELECTRONICS, INC. 5 12/30/2011
MW4024-760F-NC-WH ALLIED ELECTRONICS, INC. 1 12/30/2011